Determination of arsenic by cathodic stripping voltammetry: Sensitivity improvement using high concentration of hydrochloric acid and high temperature
Yükleniyor...
Dosyalar
Tarih
Yazarlar
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Parlar Scientific Publications (P S P)
Erişim Hakkı
info:eu-repo/semantics/openAccess
Özet
A sensitivity improvement for arsenic determination in the presence of copper by cathodic stripping voltammetry, using a high concentration of hydrochloric acid (3 mol/L) and high cell temperature (40 degrees C), is described. The optimum conditions were established, and LOD and LOQ for determination of As(III) were calculated from the calibration curve to be 7 ng/L and 22 ng/L, respectively. The accuracy of the method was checked by analyzing a certified reference material (LGC 6010-Hard Drinking Water). The method was applied to drinking water samples in Balikesir city of Turkey.
Açıklama
Anahtar Kelimeler
Arsenic, Cathodic Stripping Voltammetry, Drinking Water
Kaynak
Fresenius Environmental Bulletin
WoS Q Değeri
Scopus Q Değeri
Cilt
15
Sayı
6