The effect of screw hole inserts for the unused screw holes on the strength of a plate

dc.contributor.authorBasci, Onur
dc.contributor.authorAydemir, Selahaddin
dc.contributor.authorAcan, Ahmet Emrah
dc.contributor.authorGursan, Onur
dc.contributor.authorUzun, Bora
dc.contributor.authorErduran, Mehmet
dc.date.accessioned2025-07-03T21:25:42Z
dc.date.issued2025
dc.departmentBalıkesir Üniversitesi
dc.description.abstractIntroductionThe purpose of this study was to determine if the use of specially designed screw hole inserts in empty locking screw holes improves the strength and failure characteristics of locking plates. MethodsForty-two 7-hole locking LC/DCP plates were mounted on cylindric UHMW Polyethylene blocks with a 1-cm gap between blocks, simulating a fracture with comminution and bone loss. 21 plates had a screw hole insert placed in the center hole (centered over the simulated fracture), while 21 of the plates remained empty in the center hole. The plate-block constructs were placed in a mechanical testing machine and subjected to loading conditions. The axial, bending, and torsional stiffness and displacements needed for the failure of each plate-block construct were calculated. The statistical analysis was performed using the Mann-Whitney U test for independent variables. ResultsAll plates were then loaded to failure. There were significant differences in the axial load to failure (p = 0.017), bending load to failure (p < 0.01), and bending displacements (p < 0.01) of the test groups favoring the screw hole insert group as higher mechanical strength. Discussion/conclusionIn conclusion, the study demonstrates that the addition of the specially designed locking screw hole insert does significantly change the strength of the locking LC/DCP plates and might be suggested in the clinical application.
dc.identifier.doi10.1186/s13018-025-05686-3
dc.identifier.issn1749-799X
dc.identifier.issue1
dc.identifier.pmid40128861
dc.identifier.scopus2-s2.0-105000657486
dc.identifier.scopusqualityQ1
dc.identifier.urihttps://doi.org/10.1186/s13018-025-05686-3
dc.identifier.urihttps://hdl.handle.net/20.500.12462/21611
dc.identifier.volume20
dc.identifier.wosWOS:001451118000004
dc.identifier.wosqualityN/A
dc.indekslendigikaynakWeb of Science
dc.indekslendigikaynakScopus
dc.indekslendigikaynakPubMed
dc.language.isoen
dc.publisherBmc
dc.relation.ispartofJournal of Orthopaedic Surgery and Research
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/openAccess
dc.snmzKA_WOS_20250703
dc.subjectScrew hole inserts
dc.subjectBiomechanical stability
dc.subjectLocking compression plates (LC/DCP)
dc.subjectFracture treatment
dc.subjectPlate fatigue
dc.titleThe effect of screw hole inserts for the unused screw holes on the strength of a plate
dc.typeArticle

Dosyalar